FEATURES
USES
OPTIONAL FUNCTION LIST
▶Menu Interface
▶ 1.Integrating Function
1) Measuring Interface: Lxyi, Lxyp, Lxeq,t, Lxeq,T, Lxmax, Lxmin, Lxpeak、LAE, LC-A, SEL
Note: x is A,C,Z, y is F,S,I
2) Integrating time: 1s~99h59m59s, set in random
3) Measuring Interface: Simple, List, Huge, Big interface User interface
▶ 2.Statistical Analysis Function
1) Main Function: The statistical analysis, 24 hours noise monitor automatically.
2) Mainly Measure Index: Lxyp , Lxeq,0.5s , Lxeq,T , Lxymax, Lxymin , LxyeqT , SEL, Ln as minimum: 1, 5, 10, 50, 90 with 0.1 dB resolution, SD
Note: x is A,C,Z y is F,S,I n is 1~99
24h measures index: Ld , Ln , Ldn.
3) Up to 28 statistical Ln % values, two statistical analyzers each has 7 preset to L1, L5, L10, L50, L90, L95 & L99 and 7 user defined Ln values. Two statistical analyzers with independent time and frequency weight.
User Interface
▶ 3.Real-time 1/1 Oct Spectrum Function
1) Filter type: Parallel octave band filter, G10=103/10
2) Fulfills standards: IEC 61260: 1995 Class 1
3) Frequency bands: 11 Octave bands 16Hz-16kHz
4) Frequency Weighting: A, C, Z can be chosen
5) Center Frequency: 16 Hz, 31.5 Hz, 63 Hz,125 Hz, 250 Hz, 500 Hz, 1 kHz, 2 kHz, 4 kHz, 8 kHz, 16 kHz
6) Measuring Interface: List i and graph interface
7) Measuring Parameters: Lxyp , Lxeq,0.5s, Lxeq,T , , Lxymax, Lxymin , Tm, NR Note: x is A,C, Z,F0i y is F,S
8) Display content: Real-time display NR & NC values and curves in the process of measuring
NR according to ISO 1996:1971
NC according to ANSI S2.12-2008
9) Real-time Analysis Speed: 50 times/s
10) Level linear range: above 110dB User Interface
▶ 4.Real-time 1/3 OCT Spectrum Function
1) Filter type: Parallel (simultaneous) 1/3 octave band filter, G10=103/10
2) Fulfills standards: IEC 61260: 1995 Class 1
3) Frequency bands: 33 Octave bands 12.5Hz-20kHz
4) Real-time Analysis Speed: 50 times/s
5) Measuring Interface: List interface and graph interface
6) Measuring Parameters: Lxyp , Lxeq,0.5s , Lxeq,T , , Lxymax, Lxymin , Tm Note: x is A,C,Z,B,D, F0i y is F,S
7) Frequency Weighting: A, C, Z can be chosen
8) Level linear range: above 110dB User Interface
▶ 5.Real-time FFT Analysis Function
1) Line Number: 2048lines
2) Sampling Freq: 48 kHz, 24 kHz, 12 kHz, 6 kHz, 3 kHz
3) Measuring Parameters: MAX, MIN, LeqT
4) Window Functions: hanning, brinell, flat, rectangular User Interface
▶ 6.Dosimeter Function
1) Exchange rates: Q3, Q 4,Q 5,Q 6
2) Fulfills standard: IEC 61252: 2002
3) Selectable Thresholds: 40-90dB
4) Selectable Criterion: 70-90dB
5) Lock and with limited access
6) Noise dose: 0.01%-999.99%
7) Measuring Parameters: LAsp, LASMAX, LASMIN, TWA, LEX,8h, LCpeak, LZpeak, LAeq,T, LAVG, DOSE
8) Logging interval: 1min
9) Logging content: LAVG1m,LAeq1m, LCpeak, LZpeak,LASmax, LASmin
▶ 7.Data Logging Function
1) Logged data can be exported to PC
2) Logged content dependent on authorized and opening analyzer.
Logging of instantaneous data (Lp) and processed data (Leq and other indexes), frequency analysis.
3) Logging interval: 20ms~2000ms can be chosen, a 20ms as adjustable unit.
▶ 8.SD Card & Sound Recording Function
1) The SD card can be used as a memory card after installing the program. Saved files can be opened in the EXCEL directly
2) When connected to the computer via USB interface, it changes SD card into U disk
3) Record Format: 8000 samples/s@8bit,
48000 samples/s@32bit
4) File Format: ‘WAV’ including calibration information
5) Record Time: fs=48k, record time less than 1h per file fs=8k, record time less than 12h per file
6) Replay: by the meter or computer
Data is captured to the SD memory card inserted in the sound level meter
▶ 9.GPS Positioning Function
Measure longitude, latitude, altitude, movement speed which can be recorded together with the noise measu rement result.
▶ 10.Reverberation time measurement
If AWA6228+ sound level meter can measure the indoor reverberation time in the integrated impulse response method specified by ISO3382 after installed with reverberation time measurement software.
The integrated impulse response method means the method in which the decay curve of the indoor sound pressure level is obtained by backward integration of the squared value of the impulse response. The integrated impulse response method utilizes the modern digital processing technology that features easy and convenient operation. More important, the integrated impulse response method has higher reliability and repeatability than the interrupted noise method in principle. Particularly in the low frequency band (<250Hz) measurement, it has more advantages. Therefore, it is preferred that the integrated impulse response method is selected for the measurement. ISO 3382:1997 standard thinks that the precision of one value measured in integrated impulse response method is equivalent to the averaged value of ten values measured in interrupted noise method. It only needs to be furnished with impulse sound source, such as balloon, firecracker and starting gun etc. The sound source features compact size, light weight, easy operation, and high efficiency.
ACCESSORIES BASIC SUPPLIED
ACCESSORIES OPTIONAL
SPECIFICATIONS
|
Fulfills Standards |
IEC 61672 Class 1 |
|
IEC 61260 Class 1 |
|
|
IEC61252:2002 |
|
|
IEC 60651:2001 Type 1 |
|
|
IEC 60804: 2000 Type 1 |
|
|
ANSI S1.4: 1983 Type 1 |
|
|
ANSI S1.4A:1985 Type 1 |
|
|
ANSI S1.43:1997 Type 1 |
|
|
ANSI S1.25:1991 |
|
|
Microphone+ Preamplifier |
1/2” prepolarized condenser microphone + AWA14601 removable preamplifier (Sensitivity Level: -28dB) |
|
Correction Function |
Diffusion field correction in order to comply with standards ANSI S1.4 |
|
Frequency Range |
10 Hz ~ 20 kHz ± 1 dB (not including microphone) |
|
Total Measurement Range |
20dB-142dB (145dB Peak) |
|
Self-generated Noise |
<12 dB(A), 17 dB(C), 22 dB(Z) |
|
Frequency Weighting |
Parallel (simultaneous) A, C, Z, B, D and user1&2-defined weighting |
|
Time Weighting |
Parallel (simultaneous) F, S, I, Peak |
|
A/D Bits: |
24 bits |
|
Sampling Frequency |
48 kHz |
|
Calibration |
Using Sound Calibrator Class 1 model AWA6021A |
|
Correction Function |
Diffusion field correction in order to comply with standards ANSI S1.4 |
|
Delay Time |
The meter can delay 0~99s after pressing start measuring button |
|
Back Erase Function |
Elimination of undesired noise; example barking dogs, cars, doors |
|
Display |
240×320 color screen, adjustable brightness, backlight can be closed |
|
Display Resolution |
0.1 dB |
|
Low battery indication |
Symbol indicate low battery |
|
Data Storage |
> 3328 groups of integrating measuring results only. |
|
> 3328 groups of statistical results only (‘statistical 1’ and ‘statistical 2’ analysis index are same.) |
|
|
> 2663 groups of statistical results only (‘statistical 1’ and ‘statistical 2’ analysis index are different.) |
|
|
|
Mini-printerAH58F or AH40 |
|
Internal Clock |
Error less than 1 min/month |
| Output Interface |
AC Output (full scale): 1.0VAC RMS; Output Impedance: 1k Ω; Connector: φ3.5 mm stereo plug |
|
DC Output: 20mV/dB; Output Impedance: 1k Ω; Connector: DB-9 plug |
|
|
RS232 Interface: To computer for output some measurement results instantaneous values , also to mini-printer for printing |
|
|
USB Interface: available and no need device drive. |
|
|
Power Supply |
4×LR6 alkaline battery or rechargeable batteries |
|
5 V external power supply |
|
|
Battery Life |
Longest time of 30 hours continuously with 4×LR6 alkaline battery |
|
Dimensions |
260 (H) x 80 (W) x 30 (D), mm. |
|
Weight |
0.35 kg. (include batteries) |
|
EMC |
Type X |
|
Environment: |
Working Temperature: -10 ~ 50 °C |
|
Accessories |
Power adaptor, USB cable, Update&RS232 cable, windscreen, hand strap, windscreen fall prevention rubber, carrying case, USB-disk, user’s manual |
|
Bluetooth Module |
Wireless printing and Can communicate with smart phone and realize wireless control sound level meter |